MODEL: 50C6
Designed closely
to clinic, high quality, high security, high
dependability
Designed for special use in anesthesia room and ICU
Features
C6 series syringe pumps are developed according to
IEC60601-2-24, more safety and more advanced features
are available.
Application :
.Continuous
micro-infusion in the ICU and CCU
.Maintenance therapy micro-infusion and blood
transfusion in premature infants and neonates.
.Injection of hormones
.Continuous injection of anesthetics
.Injection of oxyfocics
.Injection of anticoagulants during dialysis and
extracorporeal blood circulation
.For anti cancer drugs
Technical data:
Flow
rate |
50ml
syringe: 0.1m/h-1200ml/h(0.1ml/h increments,
1ml/h increments when rate is over 1000ml/h)
30ml syringe: 0.1ml/h –600ml/h (0.1ml/h
increments)
20ml syringe: 0.1m/h-399.9ml/h (0.1ml/h
increments) |
Bolus
rate |
50ml
syringe: 1200ml/h
30ml syringe: 600ml/h
20ml syringe: 399ml/h |
Occlusion pressure |
40N,
60N(+/- 10N) |
Volumetric accuracy |
≤2%
(drive accuracy
≤1%) |
Total
volume indication |
0.1-9999ml |
Preset
limited volume |
0.1-9999ml |
Alarm |
Near
empty , Empty, syringe dislocated , syringe
plunger disengaged , malfunction, forget to
operate after power on, Occlusion, low battery,
battery volume exhausted, AC power failure,
preset limited volume is finished |
Operating condition |
Temperature: -5℃
to 40℃
Relative
humidity: 30 to 80% |
Transport and storage condition |
Temperature: -40℃
to 55℃
Relative
humidity:
≤95% |
External
dimensiions |
307(L) x
135(W)x 128(H) mm |
Net
weight |
2.1kg
(include internal battery and I.V. clamp) |
Syringe |
20/30/50ml |
Other functions
Occlusion pressure selected |
Lock the
fast key when infusing |
Anti
bolus |
Lock the
flow rate keys when infusing |
Syringe
brands selected |
Audio
and video alarms |
Automatically recognize syringe size
|
AC and
DC indicate |
Preset
limit volume |
RS232
port |
Total
volume indicate |
Warning
of wrong operation |
KVO |
500 log
|
|