MODEL: 50C6

Designed closely to clinic, high quality, high security, high dependability
Designed for special use in anesthesia room and ICU 
Features
C6 series syringe pumps are developed according to IEC60601-2-24, more safety and more advanced features are available.

Application :

.Continuous micro-infusion in the ICU and CCU
.Maintenance therapy micro-infusion and blood transfusion in premature infants and neonates.
.Injection of hormones
.Continuous injection of anesthetics
.Injection of oxyfocics
.Injection of anticoagulants during dialysis and extracorporeal blood circulation
.For anti cancer drugs

Technical data:

Flow rate

50ml syringe: 0.1m/h-1200ml/h(0.1ml/h increments, 1ml/h increments when rate is over 1000ml/h)
30ml syringe: 0.1ml/h –600ml/h (0.1ml/h increments)
20ml syringe: 0.1m/h-399.9ml/h (0.1ml/h increments)

Bolus rate

50ml syringe: 1200ml/h
30ml syringe: 600ml/h
20ml syringe: 399ml/h

Occlusion pressure

40N, 60N(+/- 10N)

Volumetric accuracy

2% (drive accuracy 1%)

Total volume indication

0.1-9999ml

Preset limited volume

0.1-9999ml

Alarm

Near empty , Empty, syringe dislocated , syringe plunger disengaged , malfunction, forget to operate after power on, Occlusion, low battery, battery volume exhausted, AC power failure, preset limited volume is finished

Operating condition

Temperature: -5 to 40

Relative humidity: 30 to 80%

Transport and storage condition

Temperature: -40 to 55

Relative humidity: ≤95%

External dimensiions

307(L) x 135(W)x 128(H) mm

Net weight

2.1kg (include internal battery and I.V. clamp)

Syringe

20/30/50ml

Other functions 

Occlusion pressure selected

Lock the fast key when infusing

Anti bolus

Lock the flow rate keys when infusing

Syringe brands selected

Audio and video alarms

Automatically recognize syringe size

AC and DC indicate

Preset limit volume

RS232 port

Total volume indicate

Warning of wrong operation

KVO

500 log

 

50C6T

50F6