NT2A:  Vital Signs Monitor
  • Bright, Easy-to-Read Large LED Display
  • Durable, Compact and Lightweight
  • Long-life Internal Rechargeable Battery for Uninterrupted Monitoring
  • Accurate and Reliable During Low Perfusion and Montion
  • Audible/Visual Three Level Alarms with Alarm Volume Control
  • 72-hour Data Memory Allowing On-screen Viewing of SpO2 and Pulse Rate Trends
  • Suitable for Adult, Pediatric and Neonate
  • SpO2 Accuracy:
    No Motion:
    Adults, Pediatrics: 70¡«100% ¡À2%
                        0¡«69% unspecified
    Neonates:           70¡«100% ¡À3%
                        0¡«69% unspecified
    Motion:
    Adults, Pediatrics: 70¡«100% ¡À3%
                        0¡«69% unspecified
    Neonates:           70¡«100% ¡À4%
                        0¡«69% unspecified
    Low Perfusion:
    Adults, Pediatrics: 70¡«100% ¡À3%
                        0¡«69% unspecified
    Neonates:           70¡«100% ¡À4%
                        0¡«69% unspecified

  • Pulse Rate Accuracy
    No Motion:
    Adults, Pediatrics, Neonates: ¡À3%
    Motion:
    Adults, Pediatrics, Neonates: ¡À5%
    Low Perfusion:
    Adults, Pediatrics, Neonates: ¡À5%

  • Displays and Indicators
    SpO2, Pulse Rate and Pulse Strength Bar
    Pulse Search Indicator
    Low Perfusion and Motion Warning Indicator
    Alarm Silence Indicator
    AC Power and Battery Charge Indicator
    Low Battery Warning Indicator
    Sensor Disconnect and Sensor off Warning Indicator

  • Alarm
    Audible/Visual Alarms for High/Low SpO2 and PR
    Audible/Visual Alarms for Low-Battery
    Alarm Limits Adjustable
    Alarm Volume Adjustable
    Pulse Search Indicator
    Alarm Silence Interval:30,60,90and120sec and OFF
    Low Perfusion and Motion Warning Indicator
    Sensor Disconnect Alarm
    Sensor off Alarm

¡¡

  • Performance Measurement Range
    SpO2:0¡«100%
    Pulse Rate: 20bpm¡«250bpm

  • Environment
    Opreation Temperature: 5¡«40¡æ
                           41~104¨H
    Humidity: ¡Ü80%
    Altitude: -390¡«5,000m(-1,280¡«16,404ft)
    Transport/Storage Temperature:
    -20¡«+70¡æ (-4¨H¡«158¨H)
    Humidity: ¡Ü95%

  • Output
    RS-232 Serial Port for Digital Output
    Nurse Call Function

  • Electrical Instrument
    Power Requirements:
    100~240VAC, 50/60Hz, 20VA
    Fuse Rating: 1.0A

  • Battery
    Internal Battery: 12V 3.6Ah¡¡
    Battery Capacity: ¡Ý 24 hours

  • Physical Characteristics
    Dimensions:
    264mm(W) x 95mm(H) x 172mm(D)
    [10.39in(W) x 3.74in(H) x 6.77in(D)
    Net Weight: 2.5kg/5.5ib
    Gross Weight: 3.5kg/7.7ib

  • Compliance
    SpO2: ISO9919:2005¡¡
    Safety Standards:
    En60601-1: 1990+A1: 1993+A2:
    1995+A13: 1996 Type BF-Applied
    IPX1 Approved
    Alarm: IEC60601-1-1-8: 2005
    EMC:
    EN60601-1-2: 2001, Group1 ClassA
    Environment: WEEE(2002/96/EC)

  • Thermal Arrary Recorder(Optional)
    Prints Graphical and Tabular Trend

¡¡